Optimization of tissue physical parameters for accurate temperature estimation from finite-element simulation of radiofrequency ablation.

 

Related Articles

Optimization of tissue physical parameters for accurate temperature estimation from finite-element simulation of radiofrequency ablation.

Phys Med Biol. 2015 Sep 9;60(19):N345-N355

Authors: Subramanian S, Mast TD

Abstract
Computational finite element models are commonly used for the simulation of radiofrequency ablation (RFA) treatments. However, the accuracy of these simulations is limited by the lack of precise knowledge of tissue parameters. In this technical note, an inverse solver based on the unscented Kalman filter (UKF) is proposed to optimize values for specific heat, thermal conductivity, and electrical conductivity resulting in accurately simulated temperature elevations. A total of 15 RFA treatments were performed on ex vivo bovine liver tissue. For each RFA treatment, 15 finite-element simulations were performed using a set of deterministically chosen tissue parameters to estimate the mean and variance of the resulting tissue ablation. The UKF was implemented as an inverse solver to recover the specific heat, thermal conductivity, and electrical conductivity corresponding to the measured area of the ablated tissue region, as determined from gross tissue histology. These tissue parameters were then employed in the finite element model to simulate the position- and time-dependent tissue temperature. Results show good agreement between simulated and measured temperature.

PMID: 26352462 [PubMed – as supplied by publisher]

http://www.ncbi.nlm.nih.gov/pubmed/26352462?dopt=Abstract